Differential Level & Pressure Transmitter

TH5011T Monocrystalline Silicon Differential Level & Pressure Transmitter

Product description

TH5011T sanitary differential pressure transmitter is used to measure the liquid level, density and pressure of liquid, gas or steam, and then convert it into 4 ~ 20mADC HART current signal output. TH5011T can also communicate with TH5011T handheld terminal or TH100 Modem, and use them for parameter setting and process monitoring.

Standard specification

The range is adjusted based on the standard zero point; the stainless steel 316L diaphragm; filling fluid is silicone oil.

Technical Parameters Table

Span and Range

Span and code table 1

Code

Span

Range

B

3kPa~6kPa

-6kPa~6kPa

C

4kPa~40kPa

-40kPa~40kPa

D

10kPa~250kPa

-100kPa~250kPa

Flange and minimum range relationship table 2

Nominal diameter

 

Minimum range

DN38/DN40

 

10kPa

DN50/DN51

 

10kPa

DN76.1

 

2kPa

The minimum range shall be the larger value of the minimum range in Table 1 and Table 2. The adjusted range shall not be less than the minimum range.

Performance and Specification

Reference accuracy of range adjustment

(Including linearity, hysteresis and repeatability from zero)

PH-B5011T-B

 

±0.075%

PH-C5011T-C

 

±0.1%

If TD> 10 (TD = maximum range / adjustment range), it is:

PH-B5011T-B

 

±(0.0075×TD)%

PH-C5011T-C

 

±(0.01×TD)%

Influence of ambient temperature (no capillary)

-25℃~65℃ Total influence

±(0.15×TD+0.05)%×Span

  

Between every 10 ℃ ± 0.08% × Span (when TD = 1)

-40℃~-25℃ to  65℃~85℃ Total influence

  

±(0.20×TD+0.05)%×Span

Over-range influence

±0.075%×Span

Stability

±0.1%×Span /3 years

Power effect

±0.001% /10V (12 ~ 42V DC),Negligible

Function

Zero setting

Zero and span can be adjusted to any value within the measurement range in the table As long as: nominal range ≥ minimum range

Output

Two wire 4~20mADC output with digital communications, linear or square root programmable. HART FSK protocol are superimposed on the 4~20mADC signal.

Output signal limit:Imin=3.9mA,Imax=20.5mA

Failure Alarm (the mode can be selected)

Low Mode (min):3.6 mA

High Mode (max):21 mA

No Mode (hold): Keep the effective value before the fault.

Note: The standard setting of failure alarm is High Mode.

Response Time

The amplifier damping constant is 0.1 sec; the time constant of the sensor and the liquid level flange is

0.2~2s, it depends on the range and range compression ratio.

The additional adjustable time constant is: 0.1 ~ 60s

Preheat time

< 15s

Working condition

Installation Condition

The straight transmitter body can be directly fixed at any position. The best condition is to make the process flange

The axis is in a vertical state, and the position deviation will produce a correctable zero offset.

The case can rotate up to 360° , and the positioning screw can fix it in any position.

The capillary component and the remote flange should only be installed in the same ambient temperature.

The minimum bending radius of the capillary tube is 75mm, and winding is strictly prohibited.

Ambient Temperature

-40℃~85℃

  

-20℃~65℃ (with LCD display, fluorine rubber sealing)

-40℃~70℃(OLED display )

  

Storage temperature / transport temperature

-50℃~85℃

  

-25℃~85℃(with LCD display)

Medium temperature

  

Filling fluid

Density(25℃)

Range of working temperature

Silicon oil

960kg/m3

-30~200℃

Vegetable oil

937kg/m3

0~250℃

Pressure limit

  

From 3.5kPa absolute pressure to rated pressure

Unilateral overload limit

  

The low-pressure side is the rated pressure of the transmitter body, and the high-pressure side is the rated force of the remote flange, which may have a correctable zero drift.

Electromagnetic Compatibility(EMC)

Look the EMC Performance Table

Explosion Protected Type

NEPSI Explosion-proof license:Ex dIICT6

NEPSI Intrinsically Safe License:Ex iaIICT4

Allowable temperature:-40℃~65℃

Supply & Load Requirements

The power supply voltage is 24V, the load is 520Ω, the calculation formula is as follows

Load R≤(Us-12V)/Imax kΩ,Imax=23mA

Power supply  15~36V DC

Load Working state: 0~1040Ω

Digital communication: 230~600Ω

Physical specifications

Material

Measuring capsule

316L stainless steel

Diaphragm

316L stainless steel/Hastelloy C

Process connection

316L stainless steel

Filling fluid

Silicon oil, vegetable oil

Transmitter housing

Aluminum with epoxy resin coat

Housing Gasket

Perbunan(NBR)

Name plate and tag

304 stainless steel

Electrical connections

M20X1.5 cable sealing buckle, the terminal is suitable for 0.5 ~ 2.5mm2 wire

Process connection

DN25/DN40/DN50/DN76.1 Sanitary interface

Weight

About 10-16.5kg

Degrees of Protection

IP67

EMC Performance Table

Items

Test items

Basic standards

Test conditions

Performance Level

1

Radiated interference (Housing

GB/T 9254-2008 Table 5

30MHz ~ 1000MHz

OK

2

Conducted interference DC power port)

GB/T 9254-2008 Table 1

0.15MHz ~ 30MHz

OK

3

Electrostatic Discharge (ESD) Immunity

GB/T 17626.2-2006

4kV(Line) 8kV(Air)

B

4

RF electromagnetic field immunity

GB/T 17626.3-2006

10V/m (80MHz ~ 1GHz)

A

5

Frequency magnetic field immunity

GB/T 17626.8-2006

30A/m

A

6

Electrical Fast Transient Burst Immunity

GB/T 17626.4-2008

2kV(5/50ns,5kHz)

B

7

Surge Immunity

GB/T 17626.5-2008

1kV(line to line) 2kV(line to ground)(1.2us/50us)

B

8

Conducted interference immunity induced by RF field

GB/T 17626.6-2008

3V (150KHz ~ 80MHz)

A

 